PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Assays/Reagents Not for Customer Distribution
34
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Calculated Assay Results
Recalculating Calculated Assay Results
The assays included in a calculated assay must be performed
during the same run for the calculation to take place. The
result is calculated with the original assay results and does not
automatically recalculate with an assay’s rerun data.
If a retest is necessary for one of the assays of a calculated
result, retest all the assays included in the calculated result.
Use the following instructions to move an assay’s rerun data to
the original data, and to perform recalculation of the ratio.
1. On the DATABASE screen, select the rerun sample (rerun
sample has the Processing Code “L”).
2. Select <Result> in the right column of the DATABASE
screen. The RESULT screen displays.
3. Display the desired rerun result.
4. Select <Select> in the right column of the RESULT screen.
5. Select <OK> on the Select Result dialog window. An “S”
displays to the left of the assay name and the rerun result
is copied over the original result for this assay.
6. Select <OK> to exit the RESULT screen.
7. On the DATABASE screen, select the original sample
result.
8. Select <ReCalc> in the right column. The Recalc dialog
window displays.
9. Select <Select Assay> then select the ratio to be
recalculated on the SELECT ASSAYS screen.
10. Select <OK>. The ratio is calculated from the rerun assay
data.

Table of Contents