AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Summary of Print Options
The following table summarizes the three print options for the
DOA assays.
NOTE: For troubleshooting purposes, Absorbance mode
is recommended.
ICT™ Configuration
I-Index—Calibrator/Control Position 42
I-Index is not currently used, but Calibrator/Control
position 42 (I-Index) must have a concentration of zero
defined for all ICT assays. If the concentrations are not
defined:
ICT buttons do not display on the SELECT ASSAYS FOR
CALIBRATION screen.
Calibration Summary report does not print the ICT Cal
Lvalues.
Print Options Description of Results
Outline Page
Configuration
Calibration Page
Configuration
Factor Mode—
Numeric Results
Positive or negative
numbers will print;
numbers are relative to
the cutoff calibrator
and do not represent
actual concentration.
Select Quantitative
Ranges
Calib Mode = Factor
Factor = 1,000
Factor Mode—
POS/NEG Results
Results will print POS
or NEG relative to the
cutoff calibrator.
Select Qualitative
Ranges
Calib Mode = Factor
Factor = 1,000
Absorbance Mode Results will print as
absorbance values;
results should be
interpreted as positive
or negative relative to
the absorbance of the
cutoff calibrator.
Select Quantitative
Ranges
Edit the L-Reference
field to 0.0
Calib Mode = Abs