System Troubleshooting Guide 175
94816-107—November 2004
Host Interface Trace Log
When transmission errors occur, the system generates an Error
Log Message. The message in the Display Area no longer
appears.
Use the following instructions to enable the Host Interface
Trace log.
1. Log On with the “extra” or “guest” level user code and
password.
2. Hold down the [Alt] key on the keyboard and press
[Print Scrn]. The CONTROL BOX dialog window
displays.
3. Select <Special> in the drop-down list box. The
c:\ prompt appears in the Display Area.
4. Type dbgmode at the c:\ prompt then press [Enter].
5. In the drop-down list box next to Online, select <Printer>.
6. Select <OK>.
7. Type exit at the c:\ prompt to return to the Main Display.
When the Host communication error “Illegal Message: Text
Error” is generated, ensure the Host is configured to resend the
sample.
Use the following instructions to disable the Host Interface
Trace log.
1. Log On with the “extra” or “guest” level user code and
password.
2. Hold down the [Alt] key on the keyboard and press
[Print Scrn]. The CONTROL BOX dialog window
displays.
3. Select <Special> in the drop-down list box. The
c:\ prompt appears in the Display Area.
4. Type dbgmode at the c:\ prompt then press [Enter].
5. In the drop-down list box next to Online, select <NONE>.
6. Select <OK>.
7. Type exit at the c:\ prompt to return to the Main Display.