EasyManua.ls Logo

Abbott AEROSET - Opiates 300; Cross Reactivity Testing Performed on the AEROSET System by SYVA

Abbott AEROSET
264 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Not for Customer Distribution Assays/Reagents
AEROSET
®
System Troubleshooting Guide 153
94816-107—November 2004
Opiates 300
Cross Reactivity Testing Performed on the AEROSET System by SYVA
Compounds listed in the following table yielded a negative
result at 300 ng/mL morphine (Opiates 300 calibrator) when
tested at the concentration listed.
Compound
SYVA Package Insert
µg/mL
Acetaminophen 1,000
N-Acetylprocainamide (NAPA) 400
Acetylsalicylic acid 1,000
Alpha-acetyl-N, N-dinormethadol
(dinor LAAM)
25
l-Alpha-acetylmethodol (LAAM) 25
Amitriptyline 500
Buprenorphine 100
Caffeine 1,000
Cimetidine 1,000
Clomipramine 2.5
Clonidine 1,000
Cotinine 100
Cyclobenzaprine 28
Desipramine 800
Diphenhydramine 1,000
Doxepin 10
2-Ethylidene-1,5-dimethyl-3,
3-diphenylpyrrolidine (EDDP)
1,000
Fluoxetine 500
Glutethimide 500
Ibuprofen 1,000
Ketamine 100
Ketorolac tromethamine 1,000
Lysergic acid diethylamide (LSD) 10 ng/mL
Naproxen 1,000
Nortriptyline 750

Table of Contents