AEROSET
®
System Troubleshooting Guide 103
94816-107—November 2004
Albumin BCG
Calibrators
Do not use Multiconstituent Calibrator CAL 1 (MCC 1) for
calibration. Use MCC 2 for both CAL 1 and 2. The AEROSET
System automatically dilutes MCC CAL 2 for CAL 1.
Observed Problem
Alkaline Phosphatase
Observed Problem
Illegal assay configuration
Probable Cause(s)
Corrective Action(s)
Diluent not configured. Ensure new assay parameters are installed.
(Diluent is now necessary for calibration; refer
to Revision 3 or greater of the package insert.)
Refer to Section 2, Installation Procedures and
Special Requirements of the AEROSET System
Operations Manual.
RL% Result Error Code
Probable Cause(s)
Corrective Action(s)
R2 reagent probe is out of alignment. Adjust the R2 reagent arm. Refer to Reagent
Arm Adjustment in the Hardware section of
this guide.