EasyManua.ls Logo

Abbott AEROSET - Cannabinoids; Cross Reactivity Testing Performed on the AEROSET System by SYVA

Abbott AEROSET
264 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Not for Customer Distribution Assays/Reagents
AEROSET
®
System Troubleshooting Guide 115
94816-107—November 2004
Cannabinoids
Cross Reactivity Testing Performed on the AEROSET System by SYVA
Compounds listed in the following table yielded a negative
result at 50 ng/mL 11-nor-
9
-THC-9-carboxylic acid (Multi
DOA Cutoff calibrator) when tested at the concentration
listed.
Compound
SYVA Package Insert
µg/mL
Acetaminophen 1,000
N-Acetylprocainamide (NAPA) 400
Alpha-acetyl-N, N-dinormethadol
(dinor LAAM)
25
l-Alpha-acetylmethodol (LAAM) 25
Buprenorphine 100
Caffeine 1,000
Cimetidine 1,000
Clomipramine 2.5
Clonidine 1,000
Codeine 500
Cotinine 100
Cyclobenzaprine 28
Desipramine 800
Diphenhydramine 1,000
Doxepin 10
2-Ethylidene-1,5-dimethyl-3,
3-diphenylpyrrolidine (EDDP)
1,000
Fluoxetine 500
Glutethimide 300
Ibuprofen 1,000
Ketamine 100
Ketorolac tromethamine 1,000
Lysergic acid diethylamide (LSD) 10 ng/mL
Naproxen 1,000
Nortriptyline 750
Phenytoin 1,000

Table of Contents