Cross Reactivity Testing Performed on the AEROSET System by SYVA
Compounds listed in the following table yielded a negative
result at 1,000 ng/mL d-methamphetamine (Multi DOA Cutoff
calibrator) when tested at the concentration listed.
Compound
SYVA Package Insert
µg/mL
Acetaminophen 1,000
Acetylsalicylic acid 1,000
Alpha-acetyl-N, N-dinormethadol
(dinor LAAM)
25
l-Alpha-acetylmethodol (LAAM) 25
Buprenorphine 100
Caffeine 1,000
Clomipramine 2.5
Clonidine 1,000
Codeine 500
Cotinine 100
Cyclobenzaprine 28
Doxepin 10
Glutethimide 500
Ketamine 100
Lysergic acid diethylamide (LSD) 10 ng/mL
Meperidine 1,000
Phenytoin 1,000
Tramadol 100
Zidovudine (AZT) 2 mg/mL
Zolpidem 100