PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Assays/Reagents Not for Customer Distribution
36
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
If any trends or sudden shifts in value of the verifiers occur,
check the assay reagent cartridge for onboard dating and
reagent expiration date. If necessary, replace the reagent
cartridge and repeat calibration.
NOTE: The Opiates 300 calibrator contains
methamphetamine. This does not interfere with the
Opiates 300 assay.
Refer to the assay-specific package inserts for additional
information on calibration, verifiers, and controls.
Assay Drug
Multi DOA
Verifier I
ng/mL
Opiates
300 Cal
ng/mL
Multi DOA
Cutoff Cal
ng/mL
Multi DOA
Verifier II
ng/mL
Amphetamine/
Methamphetamine
d-Methamphetamine 0 300 1,000 2,000
Barbiturates Secobarbital 0 200 800
Benzodiazepines Lormetazepam 0 200 1,000
Cannabinoids 11-nor-∆
9
-THC-9-
COOH
0 50 200
Cocaine Benzoylecgonine 0 300 1,000
Methadone Methadone 0 300 1,000
Opiates Morphine 0 300 2,000 4,000
Phencyclidine Phencyclidine 0 25 100
Propoxyphene Propoxyphene 0 300 1,000