AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Observed Problems—Software
=
Auto Rerun is not performed on replicate samples
Probable Cause(s)
Corrective Action(s)
System functionality. Manually order the rerun. Dil 1 or Dil 2 must
be selected manually, if required.
CLT Result Error Code falsely generated for a diluted sample
Probable Cause(s)
Corrective Action(s)
The following sample aspiration
sequence occurred:
1. A sample that requires a dilution is
aspirated from the sample cup.
2. A clotted sample is aspirated and a
CLT Result Error Code is
generated.
3. The sample probe moves to the
wash cup to attempt removal of
clot.
4. The diluted sample cannot be
dispensed into a new cuvette and
is flagged with a CLT Result Error
Code.
After the clot or clotted sample is removed
from the system, rerun the diluted sample.
<START> is selected on the RUN OPTIONS screen to begin a run, but the run is not
initiated; system status is READY
Probable Cause(s)
Corrective Action(s)
<START> was selected when the
system status was changing from
SUSPEND to READY.
1. Select <RUN> in the Action Area of the
Main Display. The RUN OPTIONS screen
displays.
2. Select <START> to initiate the run.