AEROSET
®
System Troubleshooting Guide 31
94816-107—November 2004
Example of Absorbance Mode Results Printout
Each assay value in this example would be compared to its
cutoff calibrator absorbance to determine if the result is
positive or negative.
The following are examples only and do not demonstrate
actual absorbance values.
The Barbiturates and Cannabinoids (THC50) assays in this
example would be reported as positive.
Amphetamine/Methamphetamine, Benzodiazepines, Cocaine,
Methadone, Opiates 300, Opiates 2,000, Phencyclidine, and
Propoxyphene would be reported as negative.
10/17/01 13:01 SN A7602220
SID 007
Name John Doe
POS 5 ( 2) PID
Date/Time Run 10/17/01 13:01
ERR NAME L/H RESULT ERR NAME L/H RESULT
PCP 0.3237 AmpM 0.2776
Prpx 0.3870 Coc 0.3212
Opi3 0.2341 Benz 0.2702
Meth 0.3988 Barb 0.2926
Assay
Sample
Absorbance
Cutoff
Absorbance
Result
Amphetamine/
Methamphetamine
0.2776 0.2778 Negative
Barbiturates 0.2926 0.2923 Positive
Benzodiazepines 0.2702 0.2705 Negative
Cannabinoids 50 0.3206 0.3206 Positive
Cocaine 0.3212 0.3232 Negative
Methadone 0.3988 0.3989 Negative
Opiates 300 0.2341 0.2378 Negative
Opiates 2,000 0.2781 0.2789 Negative
Phencyclidine 0.3237 0.3286 Negative
Propoxyphene 0.3870 0.3889 Negative