PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Assays/Reagents Not for Customer Distribution
126
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Cross Reactivity
The following digoxin cross reactivity data are not included in
the package insert.
NOTE: Test results were obtained with an alternate
assay configuration with the same antibody as
MULTIGENT
®
Digoxin.
Compound
Conc. of Cross
Reactant
Spiked
(ng/mL)
Digoxin
Reported
Value
(ng/mL)
% Cross
Reactivity
Acetaminophen 1,000,000 0.0 0.0%
Cortisol
(hydrocortisone)
10,000 0.0 0.0%
Cortisone 10,000 0.0 0.0%
Dehydroisoandrosterone 10,000 0.0 0.0%
D(+)-Digitoxose 10,000 0.0 0.0%
Estriol 10,000 0.0 0.0%
Furosemide 50,000 0.0 0.0%
Gitoxin 150 0.7 0.5%
Hydrochlorothiazide 100,000 0.0 0.0%
11α-Hydroxyprogesterone 10,000 0.1 0.001%
17α-Hydroxyprogesterone 10,000 0.0 0.0%
Lidocaine HCl 100,000 0.0 0.0%
Ouabain octahydrate 150 2.8 1.9%
Phenytoin 100,000 0.0 0.0%
Prednisolone 10,000 0.0 0.0%
Prednisone 10,000 0.0 0.0%
Procainamide 100,000 0.0 0.0%
Progesterone 10,000 0.0 0.0%
Propanolol HCl 100,000 0.0 0.0%
Quinidine sulfate
dihydrate
100,000 0.0 0.0%
Salicylate 1,000,000 0.0 0.0%
Secobarbital 100,000 0.0 0.0%
Testosterone 5,000 0.0 0.0%