EasyManua.ls Logo

Abbott AEROSET - Sheet

Abbott AEROSET
264 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Assays/Reagents Not for Customer Distribution
98
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Limit of Quantitation for Abbott Assays Data Sheet
Date: _____________ Analyte: ___________________________________________________________
Instrument: ___________________________________________ Serial No.: ______________________
Record the values for the low analyte sample replicates.
Calculate the limit of quantitation.
Comments: ____________________________________________________________________________
________________________________________________________________________________________
Technologist: _____________________________________________ Date: ________________________
Reviewed by: _____________________________________________ Date: _______________________
Lot Number Expiration Date
Reagent Kit
Calibrators
Replicate Value
1
2
3
4
5
6
7
8
9
10
A. Determine the mean of the replicates: Mean = __________________
B. Determine the SD of the replicates: SD = __________________
C. Determine the %CV of the replicates: %CV = __________________
Data Summary Calculated Value Expected Value
Limit of Quantitation
SD
Mean
× 100%CV =

Table of Contents