EasyManua.ls Logo

Abbott AEROSET - Propoxyphene; Cross Reactivity Testing Performed on the AEROSET System by SYVA

Abbott AEROSET
264 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Assays/Reagents Not for Customer Distribution
160
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Propoxyphene
Cross Reactivity Testing Performed on the AEROSET System by SYVA
Compounds listed in the following table yielded a negative
result at 300 ng/mL propoxyphene (Multi DOA Cutoff
calibrator) when tested at the concentration listed.
Compound
SYVA Package Insert
µg/mL
Acetaminophen 1,000
N-Acetylprocainamide (NAPA) 400
Acetylsalicylic acid 1,000
Alpha-acetyl-N, N-dinormethadol
(dinor LAAM)
25
l-Alpha-acetylmethodol (LAAM) 25
Buprenorphine 100
Cimetidine 1,000
Clomipramine 2.5
Clonidine 1,000
Cotinine 100
Diphenhydramine 1,000
Doxepin 10
Fluoxetine 500
Glutethimide 500
Ibuprofen 1,000
Ketamine 100
Ketorolac tromethamine 1,000
Lysergic acid diethylamide (LSD) 10 ng/mL
Meperidine 1,000
Naproxen 1,000
Phenytoin 1,000
Ranitidine 900
Scopolamine 500
Thioridazine 100
Tramadol 100

Table of Contents

Related product manuals