AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Rounding of Numbers
The AEROSET System allows only 7 digits, one of which may
be a decimal. When the last digit is 5 or greater, the number
rounds up. If the last digit is less than 5, the number rounds
down. For example, 1.15 rounds to 1.2 and 1.14 rounds to 1.1.
Sample Sequence/Smart Sampling
Smart Sampling Selected Smart Sampling Deselected
Sample Sequence
Defined
Only acknowledges ICT™ assays
defined in Sample Sequence.
Aspirates photometric assays using
Smart Sampling scheduling. Iron* is
always aspirated last.
Runs assays as defined in Sample
Sequence, except Iron. Then runs
assays in the order of buttons on the
ASSAYS screen (left to right). Iron* is
last, independent of button
placement.
Sample Sequence
Not Defined
Aspirates using Smart Sampling
scheduling. ICT assays are run later
in the run. Iron* is always aspirated
last.
Aspirates in the order of buttons on
the ASSAYS screen (left to right).
Iron* is last, independent of button
placement.
* Iron special handling is due to 2-cycle sample aspiration requirements.
Sample Sequence—will run first defined A-Line assay with first defined B-Line assay, etc.
Does run a B-Line assay in cuvette pair where ICT assays are run as A-Line.