232
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Verification Procedures/Adjustments/Alignments
Lamp Gain and A/D Check
Perform the following procedure to check lamp gains and A/D
reads. This will determine if the lamp should be replaced or if
additional troubleshooting is necessary.
Materials Required
None
Procedure
1. Ensure the system status is READY.
2. Remove one cuvette segment.
3. Select the Maintenance Utilities icon in the Action Area
of the Main Display. The MAINTENANCE UTILITIES
screen displays.
4. Select the Rxn Area tab. The Rxn Area page displays.
Figure 16: MAINTENANCE UTILITIES Screen, Rxn Area Page
5. Using the Rxn Crsl buttons, rotate the Reaction Carousel
to place the empty position in the optics light path.
6. Select Shutter <ON> to open the shutter.
Rxn Crsl
Shutter
Data