AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Barbiturates
Cross Reactivity Testing Performed on the AEROSET System by SYVA
Compounds listed in the following table yielded a negative
result at 200 ng/mL secobarbital (Multi DOA Cutoff calibrator)
when tested at the concentration listed.
Compound
SYVA Package Insert
µg/mL
Acetaminophen 1,000
N-Acetylprocainamide (NAPA) 400
Acetylsalicylic acid 1,000
Alpha-acetyl-N, N-dinormethadol
(dinor LAAM)
25
l-Alpha-acetylmethodol (LAAM) 25
Amitriptyline 1,000
Buprenorphine 100
Caffeine 1,000
Cimetidine 1,000
Clomipramine 2.5
Clonidine 1,000
Codeine 500
Cotinine 100
Cyclobenzaprine 28
Desipramine 800
Diphenhydramine 1,000
Doxepin 10
2-Ethylidene-1,5-dimethyl-3,
3-diphenylpyrrolidine (EDDP)
1,000
Fluoxetine 500
Ibuprofen 1,000
Ketamine 100
Ketorolac tromethamine 1,000
Lysergic acid diethylamide (LSD) 10 ng/mL
Meperidine 1,000
Naproxen 1,000