AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Qualitative Assay Configuration
Use the following instructions to configure a qualitative assay
in Absorbance mode.
1. Select <Assays> in the Information Access Area of the
Main Display.
2. Select the desired assay button. The ASSAY STATUS
screen displays.
3. Select <Config> in the right column of the screen. The
Assay Configuration screen displays.
4. The Outline page is the default view.
5. Verify Quantitative Ranges is selected. This is the default
setting.
6. Edit the L-Reference field from “-9999” to “0.0”.
7. Select the Calibration tab to access the Calibration page.
8. Select the list box icon next to the Calib Mode field. Select
<Abs> then select the return arrow to close the list.
9. Select <OK>.
Refer to Print Options, Absorbance Mode in this section for
additional information.