User Manual Error messages
USER MANUAL • X-MIND prime • (19) • 01/2020 • NXMPEN010C
97
X-Ray Controls
Code
Error description
360
RX button pressed on start-up or before exam
362
RX button released during emission
Sensor Ready
Code
Error description
370
Sensor ready lost during exposure
371
Sensor not ready
Temperature sensor
Code
Error description
500 ÷ 503
Temperature sensor reading error
Generator Board
Code
Error description
750
Generator board initialization error
751
Alarm "overvoltage kV"
752
Alarm "overload on filament" on Generator board
753
Alarm "overload anodic current"
754
Alarm "filament not OK"
755
Alarm "backup timer"
756
Alarm "PFC not OK"
757
Alarm "Brown OUT"
758
Alarm "NO X-ray"
759
Alarm "unexpected emission"
760
Alarm "NO RX button command"
761
Alarm "NO X-ray emission"
762
Bad unit status: emission flag detected unexpectedly
763
kV analog feedback out of range
764
mA analog feedback out of range
765
Filament analog feedback out of range
766
Generator board reset due to a brown out
767
Generator board reset due to low voltage detection
768
Generator board reset due to a watchdog timeout
769
Generator board reset due to a stack overflow
770
Mismatch between generator board (A2) and MCU board (A1) types (2D
/ 3D)
Keyboard
Code
Error description
850
One or more keycodes are pressed
852
Button >O< pressed during movements