EasyManua.ls Logo

Acteon X-MIND prime - Technical Characteristics of the X-MIND Prime; General Features and Electrical Specifications

Acteon X-MIND prime
118 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User Manual Technical characteristics
USER MANUAL • X-MIND prime • (19) • 01/2020 • NXMPEN010C
22
6. TECHNICAL CHARACTERISTICS
General features
Type
X-MIND prime
Manufacturer
de Götzen S.r.l.
21057 Olgiate Olona (VA) - Italy
Class
Class I with type B applied parts
according to IEC 60601-1
classification.
Protection degree
IPX0 standard device
Line voltage
99-264 V
Rated line voltage
110-240V
Line frequency
50/60Hz
Maximum line current
7A @ 110V 50/60Hz
3.5 A @ 240V 50/60Hz
Technical factors for maximum line
current
70kV, 7.1mA
Power consumption
1.1kVA @ 110V 50/60Hz
1.0kVA @ 240V 50/60Hz
Protection fuse (F1)
10 A T 250V 6.3x32 mm 10kA@125V
4 A T 250V 6.3x32 mm 200A@250V
Column protection fuse (F2)
3 A T 250V 6.3x32 mm 10kA@125V
1.6 A T 250V 6.3x32 mm 100A@250V
Line apparent resistance
0.4 Ω max (99-132 V)
0.5
Ω max (198-264 V)
Rated output voltage (kVp)
60 - 70kVp, with 2 kVp steps
Anodic current
2 - 7.1mA, with R20 scale steps (2, 2.2, 2.5,
2.8, 3.2, 3.6, 4, 4.5, 5, 5.6, 6.3, 7.1)
Additional filtration
≥ 2 mm Al eq.