User Manual Technical characteristics
USER MANUAL • X-MIND prime • (19) • 01/2020 • NXMPEN010C
35
6.4 Software
The equipment Graphical User Interface can be run with the software provided with the
machine or integrated in a third party imaging and database software that complies with
the following specifications: it has to be CE marked as medical device of class IIa and
integrate the equipment SDK according to what stated in the document PANOW3D API
programmer's guide Vn (n is the document revision), contact Acteon to have the latest
revision of the programmer's document.
6.5 X-MIND prime PC communication
The communication between X-MIND prime and computer is carried out with a LAN
connection based on a TCP/IP protocol.
In order to properly operate the unit, follow carefully the instructions reported in the
Service Manual at paragraph 7.6.
The system is provided with an Ethernet cat 5e cable in order to permit the PC
connection. In case of replacement, a cable of the same or superior category has to be
used.
If the communication between X-MIND prime and PC is not properly set problems in unit
connection causing impossibility of acquisition or loss of data causing distortion and
artefacts on the images can occur.
Note
X-MIND prime is not intended to transmit or receive information to/or from other
equipment through network/data couplings, but with the computer where the
unit GUI is activated.