3.3.1 Electromagnetic emissions
In accordance with the IEC 60601-1-2 Ed4 standard, X-MIND prime is suitable for use in
the electromagnetic environment specified below.
The customer or user of the system must ensure that it is used in the said environment.
Emissions test
Compliance
Electromagnetic environment
RF emissions
CISPR 11
Group I
X-MIND prime uses RF energy only for its
internal function. Therefore, its R.F.
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
Class A
X-MIND prime is suitable for use in all
establishments other than domestic and
those directly connected to the public low
voltage power supply network that supplies
buildings used for domestic purposes.
Harmonics emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies