10. MAINTENANCE
Note
Maintenance and inspection procedure must be performed without patient
positioned in the equipment.
This unit, like all other electrical appliances, must be used correctly and also serviced
and controlled at regular intervals. This precaution ensures safe and efficient
performance.
Regular maintenance consists of checks performed by the operator and/or by a qualified
technician.
The operator can control the following items:
Frequency
Type of check
Method
Daily
Functioning of the indicator lights
Visual inspection
Daily
Check that the cables do not show signs of
breaking or wear
Visual inspection
Daily
Check that the unit is not damaged
externally in such a way that the safety of
protection from radiation is compromised
Visual inspection
Daily
Check that there are no traces of oil on the
tube-head
Visual inspection
Daily
Check that arm movement is smooth
Practical inspection
Monthly
Integrity of equipment and labels
Visual inspection
Warning
If the operator detects irregularities or failures, he must immediately call
Technical Service.