User Manual Quality assurance program
USER MANUAL • X-MIND prime • (19) • 01/2020 • NXMPEN010C
47
7.6 LOG book
7.6.1 Image quality check
Dimension
Symmetry
Signal to noise
Acceptance range->
167 - 171 mm
2 mm
≥ 25 dB
Date
Measured value
Measured value
Measured value