User Manual Safety information
USER MANUAL • X-MIND prime • (19) • 01/2020 • NXMPEN010C
8
3.1.1 Precautions while using laser centring devices
For patient positioning, X-MIND prime uses two laser diodes with optical power on the
working surface < 1 mW.
The directive CEI-EN 60825-1 defines the laser as "any device that produces or amplifies
electromagnetic radiation in a coherent manner which includes a wave lengths from 180
nm to 1 mm by means of a stimulated emission". In reference to this directive, the lasers
present on the X-MIND prime are parts of class 2.
A laser in class 2 can be potentially dangerous if the ray is reflected into not protected
eyes by a mirror, watch, a ring etc.
The warning label below is affixed to X-MIND prime to indicate a laser in class 2 is
mounted internally and caution is advised:
Warning
Always keep the room well lit.
Do not look into the output windows of laser centring units.
Do not stare at the reflections of laser pointers.
Instruct the patient to keep his/her eyes closed as long as the laser pointers
are active.
Before starting an exam, the patient must remove earrings, glasses,
necklaces and any other item that could reflect the laser beam or be
impressed on the radiographic image.
Do not clean the openings of laser centring devices with tools that could
modify the optics. Any cleaning must only be performed by authorized
technicians.
Operations other than those indicated could cause the emission of
dangerous non-ionizing radiation.