4. CLEANING AND DISINFECTION
In order to guarantee a good level of hygiene and cleaning, it is necessary to carry out
the following procedures.
Warning
Disconnect the unit from the mains
before performing any cleaning.
Do not let water or other liquids penetrate the unit, as these could cause corrosion or
short circuits.
Use only a wet cloth and a mild detergent to
clean the painted surfaces, accessories and
connection cables and then wipe with a dry
cloth; do not use corrosive, abrasive solvents
(alcohol, benzene, trichloroethylene).
The centring bite or the bite protective sleeve must be replaced after each exam.
Thoroughly clean the chin support, resting handles and temple clamps group whenever
they are used.
The chin support, resting handles and temple clamps group should be disinfected (when
considered necessary) with a solution of 2% glutaraldehyde.
Note
To ensure a greater level of hygiene the handles of the equipment are covered
with a special antibacterial paint which, thanks to the emission of silver ions,
prevents the development of micro-organisms.