User Manual Quality assurance program
USER MANUAL • X-MIND prime • (19) • 01/2020 • NXMPEN010C
44
5. Check the window "Statistics": the S\N (db) value has to be higher than 25.
Record the test results in the log book at paragraph 7.6.1.
In case the tests fail, call technical assistance.