User Manual – General instructions for use
USER MANUAL • X-MIND prime • (19) • 01/2020 • NXMPEN010C
Warning
Since the X-ray button is a "dead man's switch", its release before the end of
the exposure, immediately stops the X-ray emission and the arm rotation.
Error 362 or Error 760 will be displayed
.
Warning
In case of exam interruption, let the patient exit the unit, follow the instruction
on the GUI then press >O< to carry out the axis reset
.
2. Once the exposure is completed, the system will rotate back. When it has completed
this movement, free the patient from the positioning device.
Note
If the exam is in "Test" mode with the patient already in position, the patient
must not be removed from the temple clasp group, to avoid having to reposition
him. Press the "Patient Entrance" key (6 - Figure 10) to return the unit to its
initial position. Now the system is ready to perform a new exam.
3. Press the "Patient Entrance" key (6 - Figure 10), the unit will move back to the
starting position.
Note
If you try to perform a new exam, before the cooling period has ended, a
message indicating the time to wait before performing a new exam will be
displayed.
The waiting time allows the anode in the X-ray tube to cool down.
Warning
After each exam, clean the chin support, the handles and the temple clasps
group thoroughly and change the disposable bite protective sleeve.
4. Follow instruction on paragraph 8.8.5 to perform image processing function.