Accessories and options 111
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
12.2.3.2 MicroPod calibration
Oridion supplies a Calibration kit (including a Calibration software CD and Calibration accessories)
that enables calibration of the module on a PC. The fabian HFO displays a Calibration message that
indicates on the host monitor CO
2
screen that calibration is required at the designated interval
(Figure 15).
Figure 15 - Calibration required
12.2.3.3 MicroPod calibration interval
The monitor should be calibrated by qualified service personnel after the first 1,200 operating hours
of use or 12 months, whichever comes first.
After 1,200 operating hours of use or 12 months, calibration should be performed every 12 months or
after 4,000 operating hours, whichever comes first.
12.2.3.4 Basic principles for choosing Microstream
®
etCO
2
Sampling
lines
When choosing Microstream CO
2
sampling lines, the following should be considered:
• The condition of the patient (ventilated or not ventilated)
• If the patient is ventilated, whether ventilation is humidified or non-humidified
• Patient’s size and weight
• The probability that the patient will switch between oral and nasal breathing
• Duration of use