6146/6156 DC Voltage/Current Generators Operation Manual
5.3 GPIB
5-8
5.3 GPIB
5.3.1 Overview
GPIB (General Purpose Interface Bus) allows external control of the 6146/6156 source value settings,
parameter settings, and reading data, making it simple to configure an automated measurement system.
As GPIB signals from the 6146/6156 are electrically isolated inside the unit from the source signal system,
the connection of external devices does not affect the source values.
The remote commands are the same as with USB.
General Specifications
Standard: IEEE-488.2
Code: ASCII code
Logic level: Logical 0 (High) +2.4 V min.
Logical 1 (Low) +0.4 V max.
Table 5-1 Interface Function
Code Function
SH1 Source Handshake capability
AH1 Acceptor handshake capability
T6 Basic Talker, unaddressed if MLA, serial poll
L3 Basic Listener, unaddressed if MTA, Listen Only
SR1 Service Request capability
RL1 Remote/Local switching capability
PP0 No Parallel Poll capability
DC1 Device Clear capability (The SDC and DCL commands can be
used.)
DT1 Device Trigger capability (The GET command can be used.)
C0 No Controller capability
E2 Using tri-state bus drivers