6146/6156 DC Voltage/Current Generators Operation Manual
5.8.3 Remote Command List
5-29
5.8.3 Remote Command List
1. The Default column shows default settings at Power ON or at factory shipment.
• The Power ON column shows the status when the power is turned ON.
• The *RST or RINI command initializes the settings to factory default.
However, the RINI command cannot initialize
*4
and the RINI or *RST command cannot ini-
tialize
*5
.
2. Note for description in the command list
• Parameter enclosed in [ ] can be omitted.
• Parameters enclosed in < > are single delimited data items.
• The Operation column shows the following:
Memory: Single scan or repeat scan in progress (or paused) in the memory
recall mode.
: Available
: Not available
Sweep: Operate ON in the sweep mode.
: Available
: Not available
( ):Available during Operate ON and Not available during
sweep in progress (or paused)
Class Item Command Description
Default Operation
Power
ON
Factory
shipment
Memory Sweep
Source
Output SBY Output OFF (Standby)
OPR Output ON (Operate)
SUS Output ON (Suspend)
SBY?
OPR?
SUS?
Query of the output status
Response: SBY, OPR or SUS
Suspend
HiZ/LoZ
SUZ0 HiZ: High resistance output
()
SUZ1 LoZ: Low resistance output
SUZ? Query of Suspend HiZ/LoZ
Response: SUZ0 or SUZ1
Source mode MD0 DC source mode
()
MD2 Sweep source mode
MD4 Memory recall mode
Switching during OPR/SUS sets SBY.
MD? Response: MD0, MD2 or MD4