103
Stir/Freeze Coil Replacement 3D2404R Replacement Instructions
8. Feed the replacement coil leads through
the access hole and solder the lead ends to
TR1 and TR2 on the terminal strip.
9. Reassemble by reversing steps 2-4.
Note: Make sure the yoke level doesn’t inter-
fere with the head cover.
10. Use the probe alignment tool and instruc-
tions to properly position the stir/freeze
wire.
325P404 Rev1
(3255 Service Manual)
Page 2 of 2