Appendix A
Model 3250 Specifications
1 Reference Conditions: 20 to 25 °C (68 to 77 °F); 40 to 60% Relative Humidity; tolerances of reference or calibration solutions excluded.
Serial Suffix A - C
Applies to Serial Suffix A - C. Refer to 3250 Design Changes for more information.
Electrical:
Voltage: 100 to 240 VAC (1.4 - 0.75 A)
Frequency: 50 or 60 Hz
Fuses: 2-Amp time delay (Type T) 250 VAC
Power: 95 Watts
Memory Backup: integral lithium cell; 10 years min.
(not user-replaceable)
Sample Volume: 0.2 or .25 mL
Sample Capacity: single sample
Readout: 20-character digital display
Units: mOsm/Kg H
2
O
Range:
Low: 0 to 2000 mOsm/Kg H
2
O
High: 1400 to 4000 mOsm/Kg H
2
O
Resolution:
1 mOsm/Kg H
2
O
Communications: RS-232C and barcode scanner ports
Performance at Reference Conditions
1
Linearity: Less than ±0.5% from a straight line
Repeatability:
0 to 400 mOsm ±2 mOsm/Kg (1 SD)
400 to 4000 mOsm ±0.5% (1 SD)
Drift:
Less than 1 mOsm/Kg H
2
O per month
Performance Over Operating Conditions
Temperature Effects: Less than 1 mOsm/Kg H
2
O for every 5°C (9°F) ambient
temperature change.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments Osmometer 3250 and is the answer not in the manual?

Advanced Instruments Osmometer 3250 Specifications

General IconGeneral
BrandAdvanced Instruments
ModelOsmometer 3250
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals