Appendix B
Model 4250 Specifications
1 Reference Conditions: 20 to 25°C (68 to 77°F); 40 to 60% Relative Humidity; tolerances of reference or calibration solutions excluded.
Serial Suffix A - C
Applies to Serial Suffix A - C. Refer to 4250 Design Changes for more information.
Electrical:
Voltage: 100 to 130 VAC (1.2 A) or 200 to 250 VAC (0.6 A)
Frequency: 50 or 60 Hz
Fuses: 250V time delay (Type T):
2-Amp time delay for 100-130 VAC
1-Amp time delay for 200-250 VAC
Power: 150 Watts
Memory Backup: integral lithium cell; 10 years life (typical);
(not user-replaceable)
Sample Volume: 0.2 or .25 mL
Sample Capacity: single sample
Readout: 20-character vacuum fluorescent display
Units: m °C or m °H
Range: 0 to 1000 m °C or m °H
Resolution: 1 m °C or m °H
Communications: Onboard printer, DTE EIA-232/V.24 (RS-232) serial port and optional
barcode scanner
Performance at Reference Conditions
1
Linearity: Less than ±0.5% from a straight line
Repeatability:
±2 m °C or m °H (1 SD) between 0 and -400 m °C
± 0.5% (1 SD) between -400 and -1000 m °C
Drift: Less than 1 unit per month
Performance Over Operating Conditions
Temperature Effects: Less than 1 unit for every 5°C (9°F) ambient temperature change.
Operating Conditions

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments Osmometer 3250 and is the answer not in the manual?

Advanced Instruments Osmometer 3250 Specifications

General IconGeneral
BrandAdvanced Instruments
ModelOsmometer 3250
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals