Serial Suffix D and Higher Model 4250 Specifications
Serial Suffix D and Higher
Applies to Serial Suffix D and higher. Refer to 4250 Design Changes for more information.
Electrical:
Voltage: 100 to 240 VAC (1.4 - 0.6 A)
Frequency: 50 or 60 Hz
Fuses: 2-Amp time delay (Type T) 250 VAC
Power: 95 Watts
Memory Backup: integral lithium cell; 10 years life (typical);
(not user-replaceable)
Sample Volume: 0.2 or .25 mL
Sample Capacity: single sample
Readout: 20-character vacuum fluorescent display
Units: m °C or m °H
Range: 0 to 100 m °C or m °H
High: 1400 to 4000 mOsm/Kg H
2
O
Resolution: 1 m °C or m °H
Communications: Onboard printer, DTE EIA-232/V.24 (RS-232) serial port and optional
barcode scanner
Performance at Reference Conditions
2
Linearity: Less than ±0.5% from a straight line
Repeatability:
±2 m °C or m °H (1 SD) between 0 and -400 m °C
± 0.5% (1 SD) between -400 and -1000 m °C
Drift: Less than 1 unit per month
Performance Over Operating Conditions
Temperature Effects: Less than 1 unit for every 5°C (9°F) ambient temperature change.
Operating Conditions
Temperature: 18°C to 35°C (64°F to 95°F)
Room Humidity: 5 to 80% relative humidity; (non-condensing)
Storage Temperature: -40 °C to +45 °C (-40 °F to +113 °F)
Start-up Time: From stand-by: immediately
From power on: under 5 minutes
Test Time: Approximately 90 seconds in 30-second timed mode.
2 Reference Conditions: 20 to 25°C (68 to 77°F); 40 to 60% Relative Humidity; tolerances of reference or calibration solutions excluded.