Load Resistor Replacement Instruction Advanced
Osmometer 3250/Cryoscope 4250 Service Manual
42P150R Rev1
(4D35/MK05/3255 Service Manuals)
Page 2 of 2
connectors from the 425150 load resis-
tor assembly.
b. With load resistor present, in either
location:
Unplug the existing load resistor con-
nectors, remove the mounting screws,
then remove the existing load resistor
assembly, Install the new 425150 load
resistor assembly and reconnect to the
mating connectors of the main harness
and cooling assembly.
5. Replace the instrument cover and restore
power.
6. Verify correct operation using the A/D
Test” in TEST utilities.