120 1100 Series DAD and MWD User Manual
8Maintenance
3
When replacing the Vis-lamp, assure that the Vis-lamp is
inserted as shown (flat edge towards the deuterium lamp).
4 Insert the lamp. Fix the screws and reconnect the lamp to
connector.
5 Replace the front cover. Next steps:
6
Reset the lamp counter as described in the user interface
documentation (lamps with I.D. tag cannot be reset).
7 Turn the lamp on and give the lamp 10 minutes to warm up.
8 Perform a “Wavelength Verification and Recalibration” as
described in the Service Manual (after UV-lamp
replacement).
9 Perform an “Intensity Test” as described in the Service
Manual.