128 1100 Series DAD and MWD User Manual
8Maintenance
Next steps:
9
Reconnect the capillaries, see “Exchanging a Flow Cell” on page 121.
10 Perform a leak test.
11 Insert the flow cell.
12 Replace the front cover.
13 Perform a “Wavelength Verification and Recalibration” or a “Holmium Oxide Test” as described in the Service Manual to
check the correct positioning of the flow cell.