1100 Series DAD and MWD User Manual 17
Site Requirements and Specifications 1
Analog outputs Recorder/integrator: 100mV or 1V, output range
0.001 – 2 AU, two outputs
Communications Controller-area network (CAN), RS-232C, APG
Remote: ready, start, stop and shut-down signals,
LAN
Safety and maintenance Extensive diagnostics, error detection and display
(through control module and ChemStation), leak
detection, safe leak handling, leak output signal
for shutdown of pumping system. Low voltages in
major maintenance areas.
GLP features Early maintenance feedback (EMF) for continuous
tracking of instrument usage in terms of lamp
burn time with user-setable limits and feedback
messages. Electronic records of maintenance and
errors. Verification of wavelength accuracy with
built-in holmium oxide filter.
Housing All materials recyclable.
Table 2 Performance Specifications Agilent 1100 Series DAD and MWD, continued
Type Specification Comments
NOTE
ASTM: “Standard Practice for Variable Wavelength Photometric Detectors Used in Liquid
Chromatography”.
Reference conditions: cell path length 10 mm, response time 2 s, flow 1 ml/min LC-grade
Methanol, slit width 4 nm.
Linearity measured with caffeine at 265 nm.
For environmental conditions refer to “Environment” on page 13.