1100 Series DAD and MWD User Manual 187
Control Module Screens 10
EMF (Early Maintenance Feedback)
Use the F1 key (EMF) to set EMF parameters. Choose menu item 1 (Setup
limits) to select lamp-on level for the UV and Vis lamp at which you want to
receive a warning.
If a set limit has been exceeded, a message box will pop up. This will not stop a
sequence or run (information only to plan maintenance activities). If you
press Reset, the limits will be removed. Ignore will continue to keep the EMF
flag set.