68 1100 Series DAD and MWD User Manual
4Using the Detector
Figure 37 Detector parameter (default)
up to 8 signals (A to H) with individual
wavelength settings can be selected.
spectrum settings, see page 80.
stop and post time can be set (if
required)
depending on the application, the lamps
can be selected (one or both).
peak width depends on the peaks in the
chromatogram, see page 82.
autobalance to zero absorbance (on the
analog output plus offset) at begin
and/or end of run.
mechanical slit width can be changed
for further optimization, see page 83.
margin for negative absorbance, see
page 84.
Under More additional diagnostic
signals can be added for troubleshooting
purpose, see section “Diagnostic
Signals“ in the Service Manual.
time table for programmable actions
during the run.
NOTE: The Agilent G1315C DAD and
G1365C MWD time table can contain a
maximum of 60 rows.