68 1100 Series DAD and MWD User Manual
4Using the Detector
Figure 37 Detector parameter (default)
• up to 8 signals (A to H) with individual
wavelength settings can be selected.
• spectrum settings, see page 80.
• stop and post time can be set (if
required)
• depending on the application, the lamps
can be selected (one or both).
• peak width depends on the peaks in the
chromatogram, see page 82.
• autobalance to zero absorbance (on the
analog output plus offset) at begin
and/or end of run.
• mechanical slit width can be changed
for further optimization, see page 83.
• margin for negative absorbance, see
page 84.
• Under More additional diagnostic
signals can be added for troubleshooting
purpose, see section “Diagnostic
Signals“ in the Service Manual.
• time table for programmable actions
during the run.
NOTE: The Agilent G1315C DAD and
G1365C MWD time table can contain a
maximum of 60 rows.