EasyManua.ls Logo

Agilent Technologies 1100 Series - Page 90

Agilent Technologies 1100 Series
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
90 1100 Series DAD and MWD User Manual
5 How to optimize the Detector
5 Setting the slit width
Use 4 nm slit for normal applications.
Use narrow slit (e.g 1 nm) if your analytes have narrow absorbance
bands and for high concentrations.
Use a wide slit (e.g. 16 nm) to detect very low concentrations.
spectral resolution, sensitivity and
linearity.
Optimizing spectral acquisition (DAD only)
Select spectra acquisition mode according to your needs (see
“Spectrum Settings” on page 81).
Set the spectral wavelength range (for colorless samples 190...400 nm
is sufficient).
Set step to 4 nm for normal use; set small step (and slit width) if high
resolution of spectra with fine structure is wanted.
Table 17 Optimization Overview, continued
Parameter Impact
Figure 51 Choosing a Flow Cell
Typical column
length
Typical peak
width
Recommended flow cell
T<= 5 cm
10 cm
20 cm
>= 40 cm
0.025 min
0.05 min
0.1 min
0.2 min
Typical flow
rate
Internal column diameter
0.01 ... 0.2
ml/min
0.2 ... 0.4
ml/min
0.4 ... 0.4
ml/min
1 ... 2 ml/min
0.5 ... 1 mm 2.1 mm 3.0 mm 4.6 mm
High pressure
flow cell for
pressures
above 100 bar
0.01 ... 5
ml/min
Nano
flow cell
Semi-mirco
flow cell
Standard
flow cell

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Related product manuals