Agilent 1260 Infinity DAD and MWD User Manual 185
Maintenance
9
Replacing Capillaries on a Standard Flow Cell
7
Take the new inlet capillary and bend it 90° about 35 mm
from its end.
8
Bend the capillary again by 90° as shown below.
9
Insert the capillary into the hole between fixing screw
and the inlet fitting.
10
The capillary lays in the grove and should be tied around
the body (in the grove) 5 times.
PP

