Agilent 1260 Infinity DAD and MWD User Manual 267
LAN Configuration
13
Initialization mode selection
Using Stored
When initialization mode Using Stored is selected, the parameters are taken
from the non-volatile memory of the module. The TCP/IP connection will be
established using these parameters. The parameters were configured
previously by one of the described methods.
Figure 89 Using Stored (Principle)
Using Default
When Using Default is selected, the factory default parameters are taken
instead. These parameters enable a TCP/IP connection to the LAN interface
without further configuration, see Table 30 on page 267.
Figure 90 Using Default (Principle)
1RQ9RODWLOH
5$0
6WRUHG
3DUDPHWHU
$FWLYH
3DUDPHWHU
$FWLYH
3DUDPHWHU
'HIDXOW
3DUDPHWHU
Using the default address in your local area network may result in network problems. Take
care and change it to a valid address immediately.
Tabl e 3 0 Using Default Parameters
IP address: 192.168.254.11
Subnet Mask: 255.255.255.0
Default Gateway not specified