250 Volatiles Interface
Replacement procedures
16 of 30
Jun 2001
Inlets
Agilent 6890 Gas Chromatograph Service Manual
14. Remove the screw at the top of the module as shown using a T-20 Torx
screw driver.
Figure 250-13 Removing the EPC inlet mounting screw.
15. Unlock the folded ribbon cable and remove from the connector.
Figure 250-14 The ribbon cable
Remove screw
Folded ribbon
cable