Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Faults 850
Fault 235Aux 2 thermal shutdown
Fault 235Aux 2 thermal shutdown
Status message: Aux 2 temp too hot
The aux zones temperature is higher than the maximum allowable
temperature (detector type maximum temperature plus an additional 25°C).
This indicates a defective heater drive on the mainboard or a defective sensor.
Power is disabled to the small zones and the oven. This fault can only be
cleared by turning off all the heated zones and turning the instrument off and
on again.
The problem could be caused by a defective sensor or defective heater
driver on the mainboard.
See Fault 221Thermal shutdown (status message).
Fault 236Aux 2 thermal shutdown
Status message: Aux 2 temp sensor
Aux zone 2s temperature reading reports that its temperature is lower than
expected, which indicates a shorted sensor. Power will be turned off for the
zone.
The sensor must be replaced.
See Fault 221Thermal shutdown (status message).
Fault 237No line interrupt. Thermal shutdown.
Status message: No line interrupt
The electronics on the main board are not functioning correctly. Power will
be shut off to the oven and other heated zones. The 6890 GC will not be ready.
The main board must be replaced.
See Fault 221Thermal shutdown (status message).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 Plus and is the answer not in the manual?

Agilent Technologies 6890 Plus Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 Plus
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals