850 Faults
Fault 232Back inlet thermal shutdown
12 of 22
Jun 2001
Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Fault 232Back inlet thermal shutdown
Status message: B inl temp sensor
The back inlet temperature reading reports the inlets temperature is lower
than expected, which indicates a shorted sensor. Power will be turned off for
inlet.
The sensor must be replaced.
See Fault 221Thermal shutdown (status message).
Fault 233Aux 1 thermal shutdown
Status message: Aux l temp too hot
The aux zones temperature is higher than the maximum allowable
temperature (detector type maximum temperature plus an additional 25°C).
This indicates a defective heater drive on the mainboard or a defective sensor.
Power is disabled to the small zones and the oven. This fault can only be
cleared by turning off all the heated zones and turning the instrument off and
on again.
The problem could be caused by a defective sensor or defective heater
driver on the mainboard.
See Fault 221Thermal shutdown (status message).
Fault 234Aux zone 1 thermal shutdown
Status message: Aux l temp sensor
Aux zone 1s temperature reading reports that its temperature is lower than
expected, which indicates a shorted sensor. Power will be turned off for the
zone.
The sensor must be replaced.
See Fault 221Thermal shutdown (status message).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 Plus and is the answer not in the manual?

Agilent Technologies 6890 Plus Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 Plus
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals