260 Maintaining Your GC
13 Maintaining the G3435A/G3436A FPD
4
Disconnect the retaining spring that holds the PMT assembly
to the bracket. With a rotating motion, pull the assembly
away from the filter housing.
5 To prevent light from damaging the PMT, cap the end or place
it face down.
6 Place a clean cloth under the filter housing to catch the filter.
For phosphorus filter, use the sharpened wooden tip of a
toothpick or cotton swab to dislodge the filter from the
housing.
For sulfur filter (shown below), use the wooden tip of the
cotton swab to remove the filter spacer. Then dislodge the
filter from the housing.