Maintaining the G3435A/G3436A FPD 13
Maintaining Your GC 261
7 Clean the new filter with lens tissue.
8 Install the filter in the filter housing. Install the sulfur filter
spacer, if necessary.
9 Replace the PMT assembly and secure with the spring.
10 Restore the analytical method.
Sulfur filter
Filter spacer
CAUTION
Do not use cleaning fluids. Cleaning fluids will damage lens
coatings.
CAUTION
Filters are designed for the light of the flame to pass through in a
specific direction. The triangle (on the edge of the phosphorus filter)
and the arrow (on the edge of the sulfur filter) should face away
from the flame and toward the PMT.