Installing the GC 1
Agilent 7890B Installation 57
6
Scroll to the IP address entry for the headspace sampler,
either Fr. Headspace or Bk. Headspace. Enter the headspace
sampler's IP address.
7 Press [Front Injector] or [Back Injector] as applicable.
8 A new Communication line appears. Scroll to it and press
[On/Yes]. The line changes to read Connected time. Scroll down
to display a read-only summary of the headspace sampler
configuration which the GC obtained from the sampler.
If communications fails, check the IP addresses. Verify that the
GC's IP address is correct in the headspace sampler, and that
the headspace sampler's IP address is correct in the GC. Also
check that both instruments are turned on and connected to the
LAN. Verify all LAN cable connections.
After GC installation, complete HS configuration. Refer to the
HS control software help and the HS manuals for information
on the configuration options.