72 Agilent 7890B Installation
A Making Swagelok Connections
Making Swagelok Connections
Objective
To make a tubing connection that does not leak and that can be
taken apart without damaging the fitting.
Materials needed:
1/8-inch (or 1/4-inch, if used) preconditioned copper tubing
1/8-inch (or 1/4-inch, if used) Swagelok nuts
Front and back ferrules
Two 7/16-inch (for 1/8-inch nuts) or 9/16-inch (for 1/4-inch
nuts) wrenches
1 Place a Swagelok nut, back ferrule, and front ferrule to the
tubing as shown in Figure 17.
Figure 17 Swagelok nuts and ferrules
2 Clamp a stainless steel plug or similar fitting in a bench vise.
3 Push the tubing into the stainless steel plug (see Figure 18).
Front ferrule
Back ferrule
Nut
Tubing
Important!
The narrow end of the
back ferrule fits into
the rear of the front
ferrule.
CAUTION
Do not
Use a separate stainless steel fitting in a vise for initial tightening
of the nut. Do not use an inlet or detector fitting. Strong forces are
required to properly set the ferrules, and damage to an inlet or
detector fitting is very costly to repair.

Table of Contents

Other manuals for Agilent Technologies 7890B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890B and is the answer not in the manual?

Agilent Technologies 7890B Specifications

General IconGeneral
BrandAgilent Technologies
Model7890B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals